Microbial Growth in Aqueous Mobile Phases: Troubleshooting, Prevention, and System Recovery in HPLC
for
A comprehensive guide to identifying, eliminating, and preventing microbial contamination in liquid chromatography systems using aqueous mobile phases.
Overview
Microbial contamination can develop in liquid chromatography systems whenever predominantly aqueous mobile phases are used, particularly under ambient light and room-temperature conditions. Microorganisms can colonize solvent reservoirs, tubing, pumps, mixers, degassers, autosamplers, and columns. Cells and biofilms shed both particulate matter and soluble metabolites that enter the flow path. These materials are frequently smaller than typical column inlet frit pore sizes and can pass through, accumulate within the packed bed, and alter chromatographic performance.
Once microbial growth is established, contamination often becomes system-wide, making remediation difficult and time-consuming. The resulting effects include ghost peaks, increased backpressure, retention-time instability, poor reproducibility, and compromised pump and valve performance. For this reason, prevention is considerably more effective than corrective action.
Key Symptoms and Analytical Impact
Unexpected Signals and Baseline Behavior
  • Late-eluting, broad, or variable-intensity peaks observed during blank or solvent-only injections
  • Baseline drift or instability during isocratic operation
  • Spurious features appearing during gradient runs
Pressure-Related Changes
  • Gradual increase in system backpressure over days to weeks
  • Occasional abrupt pressure spikes when biofilm fragments or particulates migrate through the system
Retention and Selectivity Effects
  • Variable retention times due to fouling at the column inlet or within the stationary phase
  • Changes in selectivity caused by adsorption of microbial byproducts
Pump and Valve Performance Issues
  • Sticking or delayed response of inlet and outlet check valves
  • Increased pressure ripple and reduced priming stability
Detector-Specific Effects
  • Elevated UV baseline and stray absorbance caused by microbial chromophores or metabolic byproducts
  • Increased chemical background, adduct formation, or ion suppression in mass spectrometric detection
Root Causes and Risk Factors
Use of mobile phases containing very high water content, particularly without organic modifiers
Exposure of clear solvent reservoirs to ambient or laboratory lighting
Extended idle periods with water-filled reservoirs and solvent lines
Poor reservoir hygiene, including topping off old solutions or reusing aged mobile phases
Insufficient filtration of aqueous solvents
Predominantly aqueous autosampler wash solutions
Introduction of microbial inoculum through samples or sample containers
Diagnostic Workflow
01
Establish a Clean Reference Baseline
Prepare fresh aqueous mobile phase and a second version containing a small percentage of organic solvent. Filter both through a fine membrane appropriate for aqueous solutions. Run blank gradients without injection and compare baseline behavior between the two solvents.
02
Solvent-Only and Zero-Volume Injection Checks
Inject solvent blanks or perform zero-volume injections. Persistent peaks in the absence of sample strongly suggest contamination upstream of the column.
03
Visual Inspection
Inspect reservoirs, caps, solvent pickup frits, and tubing for turbidity, discoloration, floating films, or settled material. Biofilms may appear as slimy residues on internal surfaces.
04
Column Bypass Test
Remove the analytical column and install a union or restrictor. Repeat blank runs. If symptoms disappear, the column or guard is implicated. If not, contamination is likely present in pumps, mixers, degassers, or the autosampler.
05
Pump Priming and Pressure Stability
Prime pumps with a mixed aqueous–organic solvent and observe pressure traces. Persistent pulsation or slow priming indicates fouled check valves or compromised degasser channels.
06
Autosampler Wash Evaluation
Review needle wash composition. Predominantly aqueous washes commonly harbor microbial growth. Test a wash containing a higher fraction of organic solvent.
07
Inline Filter Assessment
If inline or pre-column filters are installed, inspect and replace them. Accumulated particulates indicate upstream shedding from contaminated components.
Corrective Actions: System Decontamination

Important: Perform all cleaning with the analytical column removed and a bypass union installed. Observe manufacturer limits for solvents, flow rates, and materials compatibility.
A. Flow Path Flushing
  • Discard contaminated mobile phases and replace with freshly prepared, filtered solvents
  • Flush degasser and pump sequentially at moderate flow:
  1. High-purity water to dissolve soluble residues
  1. Organic-rich solvent to disrupt biofilms and remove hydrophobic material
  1. Water to remove residual organic solvent
  • Cycle mixer and proportioning valve compositions to flush all channels
  • Flush autosampler injection valve, sample loop, and wash lines with organic-rich solvent
B. Component-Level Maintenance
  • Clean solvent reservoirs and caps using laboratory detergent followed by thorough rinsing
  • Replace or clean solvent pickup frits showing discoloration or fouling
  • Install or replace inline filters to capture particulates
  • Clean or service check valves if pressure ripple persists
  • Flush degasser channels with compatible solvents within manufacturer specifications
C. Column and Guard Handling
  • Replace guard columns if backpressure increase or ghost peaks persist
  • For the analytical column:
  • Flush forward with strong organic solvent at reduced flow
  • Add a small fraction of isopropanol if compatible to disrupt biofilms
  • Avoid backflushing unless explicitly permitted by the manufacturer
  • Re-equilibrate thoroughly and reassess performance
Preventive Practices
Mobile Phase Composition
  • Include a small fraction of organic solvent in aqueous mobile phases to reduce microbial viability
  • Adjust retention and gradients as needed to accommodate composition changes
Light and Storage Control
  • Use amber reservoirs for aqueous phases
  • Minimize light exposure and heat
Solvent Preparation and Handling
  • Prepare mobile phases fresh and avoid topping off old solutions
  • Filter aqueous phases through fine membranes
  • Use high-purity water suitable for chromatographic applications
Idle System Management
  • Do not leave systems filled with pure water when idle
  • Flush with organic-rich solvent prior to shutdown or maintain slow flow
Autosampler Hygiene
  • Maintain wash solvents with sufficient organic content
  • Regularly clean needle seats and rinse ports
Inline Protection and Housekeeping
  • Use pre-column filters and replace them routinely
  • Clean reservoirs and caps on a scheduled basis
  • Avoid cross-contamination by dedicating bottles and handling them carefully
Verification After Remediation
Run solvent-only blanks to confirm baseline cleanliness
Inject mobile-phase blanks to verify absence of ghost peaks
Perform gradient tests to confirm stable backpressure and retention
Record baseline noise, pressure, and retention metrics for future comparison
Common Pitfalls and Cautions
Biocides may damage system components or interfere with detection and should not be used unless explicitly approved
Chlorine-containing cleaners and strong oxidizers are incompatible with stainless-steel flow paths
Unauthorized backflushing can permanently damage columns
Inline filters reduce risk but do not fully prevent microbial deposition within the column bed
Summary
Prevention is Key
Microbial contamination in aqueous mobile phases can severely compromise chromatographic performance and system reliability. The problem often propagates throughout the instrument and is difficult to eliminate once established.

Effective diagnosis relies on blank testing, visual inspection, column bypass experiments, and assessment of pump behavior. Remediation requires staged flushing with organic-rich solvents, replacement of sacrificial components, and careful column handling. Long-term control depends on modifying mobile phase composition, protecting solvents from light, maintaining reservoir hygiene, and avoiding prolonged idle periods with water-filled systems.